8 – Language Dictionary
TRIGger:SEQuence2:LEVel:CURRent
TRIGger:ACQuire:LEVel:CURRent
This command sets the trigger level for internally triggered current measurements. A positive current
trigger occurs when the current level changes from a value less than the lower hysteresis band limit to a
value greater than the upper hysteresis band limit. Similarly, a negative current trigger occurs when the
current level changes from a value greater than the upper hysteresis band limit to a value less than the
lower hysteresis band limit.
Command Syntax TRIGger:SEQuence2:LEVel:CURRent<NRf+>
Parameters 0 to MAX (see table 8-3)
*RST Value 0
Examples TRIG:SEQ2:LEV:CURR 5
Query Syntax TRIGger:SEQuence2:LEVel:CURRent?
Returned Parameters <NR3>
Related Commands TRIG:SEQ2:LEV:VOLT
TRIGger:SEQuence2:LEVel:VOLTage
TRIGger:ACQuire:LEVel:VOLTage
This command sets the trigger level for internally triggered voltage measurements. A positive voltage
trigger occurs when the voltage level changes from a value less than the lower hysteresis band limit to a
value greater than the upper hysteresis band limit. Similarly, a negative voltage trigger occurs when the
voltage level changes from a value greater than the upper hysteresis band limit to a value less than the
lower hysteresis band limit.
Command Syntax TRIGger:SEQuence2:LEVel:VOLTage<NRf+>
Parameters 0 to MAX (see table 8-3)
*RST Value 0
Examples TRIG:SEQ2:LEV:VOLT 5
Query Syntax TRIGger:SEQuence2:LEVel:VOLTage?
Returned Parameters <NR3>
Related Commands TRIG:SEQ2:LEV:CURR
110
TRIGger:ACQuire:LEVel:CURRent<NRf+>
Unit A (amperes)
TRIG:ACQ:LEV 2
TRIGger:ACQuire:LEVel:CURRent?
TRIGger:ACQuire:LEVel:VOLTage<NRf+>
Unit V (volts)
MAX
TRIG:ACQ:LEV 2
TRIGger:ACQuire:LEVel:VOLTage?
TRIG:ACQ:LEV:CURR MAX
TRIG:SEQ2:HYST:CURR
TRIG:ACQ:LEV:VOLT
TRIG:SEQ2:HYST:VOLT