E – Line Voltage Conversion
white/red/grey
orange
orange
grey
grey
orange
orange
white/red/grey
Figure E-1, Power Transformer AC Input Connections
Install the Correct Line Fuse
Unscrew the line fuse cap from the rear panel and install the correct fuse.
♦
For 100/120 Vac operation: 2 A slow-blow; HP part number 2110-0303
For 220/230 Vac operation: 1 A slow-blow; HP part number 2110-0007
Mark the voltage setting that the unit has been set to on the rear panel label.
♦
Close the Unit
Replace the outer cover.
♦
♦
Reconnect the power cord and turn on the unit.
148
grey
orange
white/yellow
white/violet
white/red/grey
grey
white/yellow
white/violet
white/red/grey
white/red
white/black
white/brown
orange
white/red
red
white/black
black
white/brown
orange
(spare)
white/yellow
white/violet
orange
(spare)
white/yellow
white/violet