Metering Keys

Metering keys control the metering functions of the dc source. As set from the factory, all front panel
measurements from the main output (output 1), are calculated from a total of 2048 readings taken at a
46.8 microsecond sampling rate. Therefore, the factory default acquisition time for a single front panel
measurement is about 100 milliseconds. Refer to "Making Front Panel Measurements" for more
information about changing the front panel sampling rate and the number of measurement points.
All front panel measurements from the DVM and from output2 are fixed at 2048 measurement readings
taken at a 15.6 microsecond sampling rate.
NOTE:
The front panel sample rate and data point settings are separate and independent of the
sample rate and data point settings that are programmed over the GPIB interface. When
an GPIB measurement is in progress, the front panel display temporarily indicates
-- -- -- -- --. Front panel measurements resume when the GPIB measurement completes.
Meter
Shift
Input
Notes:
Press this key to access the meter menu list. Also use this key to exit a
menu at any time and return to meter mode.
Display
<reading>V <reading>A
<reading>V MAX
<reading>V MIN
<reading>V HIGH
<reading>V LOW
<reading>V RMS
<reading>A MAX
<reading>A MIN
<reading>A HIGH
<reading>A LOW
<reading>A RMS
<reading>V DC:DVM
<reading>V RMS:DVM
Press this key to access the following metering functions.
Display
CURR:RANGE <char>
CURR:DET <char>
TINT <value>
POINTS <char>
1
only valid for Agilent 66321D/66319D
reading = the returned measurement
value = a numeric value
char = a character string parameter
'
(
Use
and
to scroll through the menu commands.
Use
#
and
&
to scroll through the menu parameters.
Use
and
to select a digit in a numeric entry field.
"
!
Measurement
Measures output dc voltage and current
Measures peak output voltage
Measures minimum output voltage
Measures the high level of a voltage waveform
Measures the low level of a voltage waveform
Measures rms voltage
Measures peak output current
Measures minimum output current
Measures the high level of a current waveform
Measures the low level of a current waveform
Measures rms current
Measures dc voltage on DVM input
Measures rms voltage on DVM input
Command Function
Select current range (3A | 1A | 0.02A | AUTO)
Select current measurement bandwidth
Sets the front panel measurement interval in seconds
(15.6 µs to 1 second)
Sets the # of points in front panel measurement buffer
( 1, 2, 4, 8, 16, 32, 64, 128, 256, 512, 1024, 2048)
Front Panel Operation - 5
1
1
(ACDC | DC)
47