Configure the RS-232 interface as follows:
Action
1.
On the System keypad, press Address.
2.
Scroll through the Address menu by pressing
select the RS-232 interface. The baudrate command lets you select the baudrate. The
parity command lets you select the parity. The flow command selects the flow control
options.
3.
keys let you select the command parameters. Press Enter when done.
The
and

6 - Saving and Recalling Operating States

You can save up to 4 states (from location 0 to location 3) in non-volatile memory and recall them from
the front panel. All programmable settings are saved. This capability is only available when the unit is set
to the SCPI programming language.
NOTE:
You can program the unit to automatically power-on according to the operating state that
is saved in location 0 as shown in the third example on this page.
Save an operating state in location 0 as follows:
Action
1.
Set the instrument to the operating state that you want to save.
2.
Save this state to location 0. Press Save, Enter Number, 0, Enter.
Recall a saved state as follows:
Action
1.
Recall the state saved in location 0 by pressing Recall, Enter Number, 0, Enter
Select the power-on state of the dc source as follows:
Action
1.
On the Function keypad, press Output, and scroll through the Output menu until you
get to the PON state command.
2.
Use the
and
keys to select either RST or RCL0. RST sets the power-on state of
the unit as defined by the *RST command. RCL0 sets the power-on state of the unit to
the state saved in *RCL location 0. Press Enter when done.
Clear the non-volatile memory of the dc source as follows:
Action
1.
On the Function keypad, press Output, Enter. This returns the unit to the factory-
default settings.
2.
Save these settings to location 0. Press Save, Enter Number, 0, Enter.
3.
Repeat step #2 for memory locations 1 through 3.
. The interface command lets you
Front Panel Operation - 5
Display
ADDRESS 5
INTF RS232
BAUDRATE 9600
PARITY EVEN
FLOW XON-XOFF
Display
*SAV 0
Display
*RCL 0
Display
PON:STATE RST
Display
*RST
*SAV 0
*SAV 1
*SAV 2
*SAV 3
45