Trg; Tst; Wai - Agilent Technologies E4356A Operating & Programming Manual

Telecommunications dc power supply
Table of Contents

Advertisement

Query Syntax
Returned Parameters
Related Commands

*TRG

Meaning and Type
Trigger Device Trigger
Description
This command generates a trigger when the trigger subsystem has BUS selected as its source. The command has the same
affect as the Group Execute Trigger (<GET>) command.
Related Commands

*TST?

Meaning and Type
Test Device Test
Description
This query causes the power supply to do a self-test and report any errors (see "Selftest Error Messages" in "Chapter 3 -
Turn-On Checkout").
Returned Parameters

*WAI

Meaning and Type
Wait to Continue
Device Status
Description
This command instructs the power supply not to process any further commands until all pending operations are completed.
"Pending operations" are as defined under the *OPC command. *WAI can be aborted only by sending the power supply a
GPIB DCL (Device Clear) command.
*STB?
Command Syntax
Parameters
Query Syntax
Query Syntax
Related Commands
Command Syntax
Parameters
Query Syntax
Related Commands
<NR1>
(Register binary value)
(None)
*TRG
(None)
(None)
ABOR CURR:TRIG INIT TRIG[:IMM] VOLT:TRIG
<GET>
*TST?
<NR1>
0
Indicates power supply passed self-test.
Nonzero
Indicates an error code.
(None)
*WAI
(None)
(None)
*OPC *OPC?