There is an internal calibration routine (Factory Calibration) that collects
correction data for both the baseband and RF magnitude and phase errors
over the entire RF frequency and power level range on any unit with
options 653, 655, 656, and 657. The internal channel correction cannot be
turned on until after the Enhanced Factory Calibration has been executed
once. See
The internal channel correction feature flattens the system magnitude and phase response across
the maximum bandwidth supported by the instrument (up to 160 MHz BW, depending on the
option). This performance improvement is available at any arbitrary center frequency or amplitude
level.
Correction filter generation and application is performed internally. The correction filtering occurs
in real time and is applied to live data.
There are two correction types: factory and user. The Factory Calibration optimizes the performance
at the front panel RF output connector into a precision 50 ohm load. The User Calibration can
extend this performance to a new calibration reference plane and deliver the highest signal quality
to the user's DUT.
This feature also minimizes unwanted baseband images by flattening (and matching) the
magnitude and phase response of the I and Q channels from the signal DACs to the IQ modulator.
This calibration should be run when the ambient temperature has varied by at least ±5 degrees
Celsius from the ambient temperature at which the previous calibration was run.
When this feature is off, the box will behave as it always has. When this feature is on, the internal
I/Q path is active, the I/Q Correction Optimized Path is RF Output, and the frequency is changed by
more than 1 kHz, the firmware will calculate a channel correction filter. For List/Sweep, the
calculation will be done prior to the first sweep using the specified frequencies when either
waveform sweep is active or the baseband is on and the instrument is optimized for the internal
path. This calculation will cache corrections for up to the maximum number of cache points (256).
For list/sweep, the sweep will pause and recalculate the correction caches before running again.
In the case of arbitrary frequency switching, once the correction cache is
full (256 unique frequency points), the oldest frequency corrections will be
forgotten as new frequencies are selected.
Whenever I/Q Timing Skew, I/Q Delay, Quadrature Angle Adjustment, or
the Int Equalization Filter is adjusted, all caches are forgotten.
Additional characteristics of the internal channel correction:
— When the internal channel correction is on, arbitrary frequency switching while the baseband is
on will take up to an additional 290 ms (72 ms typical) the first time that frequency is specified.
After the first time that a frequency is selected, switching to that frequency again takes an
additional 1 ms.
Keysight CXG, EXG, and MXG X-Series Signal Generators User's Guide

Using Internal Channel Correction (N5166B/72B/82B Only)

"Perform Enhanced Factory Calibration" on page
Optimize Performance