Planned maintenance
SERIES 1200 LIQUID TURBINE METER
September 2020
4. Remove the complete internals of the meter.
9.3.2
Disassemble internal meter housing components - NPS
1.5 through 2
Procedure
1. Remove the nut, washer and flow conditioning plate by pulling the parts from the
60
www.Daniel.com
70
Important
The shaft and rotor assembly may come out with this. The bearings, shaft and rotor
will fall out of the meter body if tipped. Handle the rotor with care. Improper
handling of the rotor assembly may cause distortion to the rotor blades thereby
affecting meter performance.
Figure 9-3: Upstream stator removal
Figure 9-4: Internals of the meter
center bolt.