Levelflex FMP50 PROFIBUS PA
Additional information
Endress+Hauser
Meaning of the display options
• Settings identical
The current device configuration of the HistoROM is identical to the backup copy in the
display module.
• Settings not identical
The current device configuration of the HistoROM is not identical to the backup copy in
the display module.
• No backup available
There is no backup copy of the device configuration of the HistoROM in the display
module.
• Backup settings corrupt
The current device configuration of the HistoROM is corrupt or not compatible with the
backup copy in the display module.
• Check not done
The device configuration of the HistoROM has not yet been compared to the backup copy
in the display module.
• Dataset incompatible
The data sets are incompatible and can not be compared.
To start the comparison, set Configuration management (→  169) = Compare.
If the transmitter configuration has been duplicated from a different device by
Configuration management (→  169) = Duplicate, the new device configuration
in the HistoROM is only partially identical to the configuration stored in the display
module: Sensor specific properties (e.g. the mapping curve) are not duplicated. Thus,
the result of the comparison will be Settings not identical.