Hygienic reprocessing at
home

Cleaning

• Place all nebuliser handset parts as well
as the aerosol head in warm
(approx. 40 °C) tap water with a little
dishwashing soap (dosage according to
the instructions of the manufacturer) for
5 minutes.
Cleaning the aerosol head
• After immersing in water, rinse the front
and back of the aerosol head for
approx. 1 minute under warm running
tap water (approx. 40 °C).
®
eFlow
rapid - 03/12
• If parts are heavily soiled, clean the aer-
osol head again by shaking in the water
in the bowl and then rinsing with running
tap water.
Caution:
- Do not put the aerosol head in a micro-
wave oven.
- Do not attempt to clean the aerosol
head in a dishwasher.
- Mechanical cleaning of the aerosol
head with brushes or abrasives may
cause irreparable damage!
Cleaning other parts of the nebuliser
handset (not the aerosol head)
• After immersing in water, clean these
nebuliser handset parts thoroughly, if
necessary using a soft, clean toothbrush,
and then rinse off with warm (approx. 40
°C), running tap water.
You can remove excess water more
quickly by shaking all parts.
Cleaning the control unit and
connection cord
en
53