Immunity test
IEC 60601 test
level
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2,5 GHz
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) tele-
phones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic
site survey should be considered. If the measured field strength in the location in which the
®
eFlow
rapid is used exceeds the applicable RF compliance level above, the eFlow
should be observed to verify normal operation.
If abnormal performance is observed, additional measures may be necessary, such as reorient-
ing or relocating the eFlow
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
Note 1: At 80 Hz and 800 MHz, the higher frequency range applies.
Note 2: This guidance may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
®
eFlow
rapid - 03/12
Compliance
level
10 V/m
Table 2 and 4 of DIN EN 60601-1-2:2007 (IEC 60601-1-2:2007)
®
rapid.
Electromagnetic environment
– Guidance
Recommended separation
distance:
d = 0.35 * SQRT (P)
d = 0.35 * SQRT (P) 80 MHz to
800 MHz
d = 0.70* SQRT (P) 800 MHz to
2.5 GHz
Where „P" is the maximum out-
put power rating of the transmit-
ter in watts (W) according to the
transmitter manufacturer and
„d" is the recommended sepa-
ration distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site
a)
survey
, should be less than
the compliance level in each
b)
frequency range
.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
de
®
rapid
323