Separation distances
®
The eFlow
rapid is intended for use in the electromagnetic environment in which radi-
ated RF disturbances are controlled. The customer or user of the eFlow
prevent electromagnetic interference by maintaining a minimum distance between por-
table and mobile RF communication equipment (transmitters) and the eFlow
recommended below, according to the maximum output power of the communications
equipment.
Recommended separation distances for portable and mobile RF communication
equipment and the eFlow
Rated maximum
output power of
transmitter [W]
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended
separation distance „d" in meters [m] can be determined using the equation applicable
to the frequency of the transmitter, where „P" is the maximum output power rating of the
transmitter in watts [W] according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidances may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
324
®
rapid
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 0.35 * SQRT (P)
0.04
0.11
0.35
1.11
3.50
[m]
80 MHz to 800 MHz
d = 0.35 * SQRT (P)
0.04
0.11
0.35
1.11
3.50
Table 6 of DIN EN 60601-1-2:2007 (IEC 60601-1-2:2007)
®
rapid can help
®
rapid as
800 MHz to 2.5 GHz
d = 0.70 * SQRT (P)
0.07
0.22
0.70
2.21
7.00
®
eFlow