Troubleshooting - Pari eflowrapid 178G1005 Instructions For Use Manual

Nebuliser system
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 39

TROUBLESHOOTING

7
Faults in installation or operation (e.g. neb-
uliser handset not connected, no medica-
tion added, power supply interrupted) are
displayed approx. 5 to 6 times in flashing
mode. The device then switches off. If sev-
eral causes of malfunction exist, they will be
Display indicator
(flashes alternately)
58
displayed in the following order each time
the device is switched on again. Please fol-
low the step-by-step instructions for cor-
recting such malfunctions (see following
list):
Possible cause
Battery flat.
Cable connection to the
control unit or on the
nebuliser handset inter-
rupted.
No medication added.
Inhalation interrupted.
Battery charge level
50%.
The maximum operat-
ing time of 20 min. for
one application has
been exceeded.
Solution
Insert new batteries or
operate with the power
adapter connected.
Check cable
connections.
Add medication.
Resume inhalation by
restarting the device
(press ON/OFF button).
Have new batteries on
hand or operate with the
power adapter
connected.
Resume inhalation by
restarting the device
(press ON/OFF button).
®
eFlow
rapid - 03/12