Electromagnetic immunity
Immunity test
IEC 60601 test
level
Electrostatic
±6 kV contact
discharge
±8 kV air
(ESD)
IEC 61000-4-2
Electrical fast
±2 kV for power
transient /burst
supply lines
IEC 61000-4-4
±1 kV for input /
output lines
Surge
±1 kV common
mode
IEC 61000-4-5
±2 kV differential
mode
Voltage dips,
<5% U
short interrup-
(>95% dip of U
tions and volt-
for 1/2 cycle
age variations
40% U
on power sup-
(60% dip of U
ply input lines
for 5 cycles
IEC 61000-4-11
70% U
(30% dip of U
for 25 cycles
<5% U
(>95% dip of U
for 5 s
Power fre-
3 A/m
quency (50 Hz/
60 Hz) mag-
netic field
IEC 61000-4-8
Conducted RF
3 V
IEC 61000-4-6
150 kHz to
80 MHz
a) U
is the mains voltage prior to application of the test level.
T
322
Compliance
level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for input /
output lines
±1 kV common
mode
±2 kV differential
mode
a)
<5% U
T
)
(>95% dip of U
T
for 1/2 cycle
40% U
T
)
(60% dip of U
T
for 5 cycles
70% U
T
)
(30% dip of U
T
for 25 cycles
<5% U
T
)
(>95% dip of U
T
for 5 s
3 A/m
10 V
eff
150 kHz to
80 MHz
Electromagnetic environment
– Guidance
Floors should be wood, con-
crete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%.
Mains power quality should be
similar to that of a typical com-
mercial or hospital environ-
ment.
Mains power quality should be
similar to that of a typical com-
mercial or hospital environ-
ment.
Mains power quality should be
T
)
similar to that of a typical com-
T
mercial or hospital environ-
ment. When the user of the
T
eFlow
)
T
also calls in the event of disrup-
tion of supply, it is recom-
mended the eFlow
T
)
an uninterruptible power supply
T
or a battery.
T
)
T
Power frequency magnetic
fields should be at levels char-
acteristic of a typical location in
a typical commercial or hospital
environment.
Portable and mobile RF com-
eff
munication equipment should
be used no closer to any part of
the eFlow
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
®
rapid continued function
®
rapid from
®
rapid, including
®
eFlow
rapid - 03/12