Description of this Upgrade
Step 1. Verify PNA and Erase Flash
"Step 21. Verify the PNA Analyzer Program is Running with the Correct
Options" on page 47
"Step 22. Perform Post-Upgrade Adjustments and Calibration" on page 48
"Step 23. Prepare the PNA for the User" on page 51
Step 1. Verify PNA and Erase Flash
This section verifies the PNA boots normally and if OK, deletes the flash in
preparation for installing and calibrating the new synthesizer assembly. Refer
to
1. Turn on your PNA.
2. Verify it boots up without errors.
3. If your PNA boots up normally.
4. Open the SCPI Parser Console: System > System Setup... > Remote
Installation Note N5222-90133
Figure 2 on page
14.
If there are errors, refer to your instrument's Chapter 4 "Troubleshooting"
in the PDF Service Guide
Interface... > then click Show SCPI Parser Console
1. See
"Downloading the Online PNA Service Guide" on page
1
.
7.
13