Description of this Upgrade
Step 22. Perform Post-Upgrade Adjustments and Calibration
3. LO Drive Adjustment
Figure 29-1
Connect sensor at the end of one of the cables from the HMA26.5
4. Perform PNA Source Power Adjustment.
If the instrument has Option 422/423, it may have Option XSB installed
and SOURCE3
Remember to reconnect the load to SR3 J13 on the rear panel.
1. Option XSB (Source 3) is only orderable on instruments with Option 422/423.
5. In the Synthesizer Calibrations – WSynth Calibrations window: Press
In the Synthesizer to Calibrate area, only the sources that are installed are
available for calibration (i.e., Source 2 and Src 3 (XSB)
(grayed out) when not installed).
1. Option XSB (Source 3) is only orderable on instruments with Option 422/423.
— 10 MHz frequency reference adjustment
— EE default adjustment: All Synthesizers (Version 7 synthesizers)
— source adjustment
— IF gain adjustment
Installation Note N5222-90133
1
will be available to adjust.
Program Files > Keysight > Network Analyzer > Service > WSynth
1
are inactive