Preparing the Instrument
Evacuating the System
Some systems are equipped with vacuum capabilities; all are equipped with purge capabilities. Purge gas
exits through the anti-siphon exit port assembly.
CAUTION: Never place any type of valve between the EM exhaust and the anti-siphon exit port
assembly.
MISE EN GARDE: Ne placez jamais un type de vanne quelconque entre l'échappement EM et
l'ensemble de l'orifice de sortie anti-siphon.
The EM is equipped with a vacuum meter assembly for determination of the vacuum level.
Vacuum and Purge Operation
1
Verify all vacuum connections.
2
Check the oil level in the anti-siphon exit port. Add vacuum pump oil as needed to reach the level marked
on the outside of the reservoir.
3
Close all purge valves.
Figure 32
4
Turn on the vacuum pump.
5
Evacuate the system to at least 50 mTorr.
DXF Getting Started Guide
Core Purge Valve
Page 49