2
From the Diagnostics menu, select Temperature Controller. A Temperature Controller window
displays.
Figure 47 Temperature Controller window.
3
Click Initialize.
4
Click System Enable both from the Temperature Controller window and on the front panel of the
instrument.
5
Click Furnace Enable both from the Temperature Controller window and on the front panel of the
instrument.
6
Enter 200 in the field next to the Setpoint button. Do not click the Setpoint button.
7
Enter 10 in the field next to Ramp Rate and click Ramp Rate.
8
Enter 120 in the field next to Ramp Dwell and click Ramp Dwell.
9
Click Ramp Setpoint to start heating.
10 After 2 hours at 200°C, the heating stops and the system cools down.
The next time that the furnace is used for a new test, the vacuum should be below 50 mTorr in less than 30
minutes.
DXF Getting Started Guide
Page 63