Conditions When Output Data Become Unfixed - Panasonic HL-G1 A-C5 Series User Manual

Compact laser displacement sensor
Hide thumbs Also See for HL-G1 A-C5 Series:
Table of Contents

Advertisement

2-5

Conditions When Output Data Become Unfixed

The operation status will be judged as a data unfixed status (i.e., there is no determined
data) immediately after settings changes are made, the system is reset, or laser stop input
is turned ON. This status is not an alarm status.
The data unfixed status starts at the time of restarting measurement after setting
refreshment or resetting until the number of data measurement times reaches an
average.
Under the data unfixed status, a digital output of -999.9999 [mm] and an analog
output of 11.000 [V] or 21.6 [mA]
*1 This is the initial value. Under the data unfixed status, the analog output of the
system can be changed to a fixed value. For more information,
"3-3-7 Alarm Settings" – "Analog Output at Alarm".
*2 Under the data unfixed status, the zero set input will be ignored.
*3 For cancelling the data unfixed status,
For example, the graph on the right-hand
side shows the digital display at reset input
for NORMAL measurement.
The table below shows the measurement
value display and output under the data
unfixed status or alarm status.
Status
Digital output
Analog output
Terminal
output
*1
will be turned ON.
Data unfixed status
Measurement data is unfixed.
(Average buffer has not reached
the setting number of times.)
-999.9999 [mm]
Interlocks with the analog output
at alarm
OFF
2-9
Chapter 2 I/O Terminal Block
refer to Chapter 5 "Troubleshooting".
Reset input
Data unfixed status
(Data remeasurement
period)
Alarm status
Measurement disabled status due to
poor light intensity or when the
target object is off the measurement
range.
The previous value (default) is kept
on hold.
The previous value (default) is kept
on hold.
OFF
2
refer to

Advertisement

Table of Contents
loading

Table of Contents