Installation

Installing the Anti-tilt Anchor

Installing the Anti-tilt Anchor
Table-top units
The anti-tilt anchor secures the top device in a stack to a solid part of a building. The anti-tilt
anchor is to be mounted on the side opposite of the door hinges.
Bend the fixing tabs of the anti-tilt anchor up on one side and down on the other by an angle
of approx. 90°.
1. Do not use this position if the door is hinged on this side. Right-hand hinges represent the
standard configuration.
2. Preferred position.
3. Alternative position. Do not use if the door is hinged on this side.
Remove the bracket screws. Use the preferred position, if possible.
Fix the anti-tilt anchor with the bracket side down to the unit (see figure).
Position the unit with the anti-tilt anchor to in an angle of approx. 90° +/- 20%.
Take care that the stacking feet of the unit are still in correct place on the lower unit or on the
stacking adapter.
Fix the anti-tilt anchor to a solid part of the building.
Floor stand units
Heratherm floor stand units must always be attached to the wall using two
(2) retaining brackets on the outer left and right side on the back of the
unit.
3-12
Heratherm
Thermo Scientific