Installing The Unit - Thermo Scientific 3110 Operating And Maintenance Manual

3100 series forma series ii water jacket co2 incubator
Hide thumbs Also See for 3110:
Table of Contents

Advertisement

Section 1
Installation and Start-Up

Installing the Unit

1-4
Water Jacketed CO
Incubator
2
RUN
SETPOINT
Default Mode Temperature
Overtemp
CO2
O2
Base Unit Displays
Option Displays
1. Locate the unit on a firm, level surface capable of supporting the unit's
operational weight of 365 lbs. (166kg) plus the estimated total sample
and user equipment weight. The overall space should be well ventilated
and large enough that the concentration of carbon dioxide (CO2)
and/or nitrogen (N2) gases cannot accumulate, thus displacing the
oxygen (O2) to dangerous levels in the room. Read Connecting CO2
or N2 (O2 System) Gas Supply below carefully.
Note If two incubators are to be stacked, then the surface must be capable
of twice the weight estimated above.
2. Locate the incubator away from exterior doors and windows as changes
in outdoor temperatures and contact with direct sunlight can affect the
anti-condensation functions of the unit.
3. Position the door opening away from forced air heating and cooling
ducts as these can carry dust, dirt, and other contaminants into the
incubator chamber, as well as negatively affect the door opening
recovery time.
CALIBRATION
CONFIGURATION
Temp Offset
Audible
CO2 Cal
New HEPA timer
IR CO2 Span
Replace HEPA reminder
O2 Cal at 20.7%
Access Code
O2 Offset
Temp Lo Alarm
RH Offset
Temp Hi Alarm
Temp Relay
CO2 Lo Alarm
CO2 Hi Alarm
CO2 Relay
T/C CO2 Z #
T/C CO2 S #
RH Lo Alarm
RH Relay
O2 Lo Alarm
O2 Hi Alarm
O2 Relay
Display 1 Setup
Display 2 Setup
Gas Guard Setup
RS485 Address
Thermo Scientific

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

312031113121313031313140 ... Show all

Table of Contents