Installing the Stacking Adapter

Stack as follows when using a stacking adapter (numbers denote incubator volumes in liters):
• for 60 l or 100 l on 180 l use stacking adapter Heratherm 180L,
• for on 100: use stacking adapter Heratherm100L.
To prevent the top device from slipping and dropping down, the following requirements must
be fulfilled before devices may be stacked:
• Only two units may be stacked together. When stacking devices with the same type of
enclosure, yet with a different footprint the device with the smaller footprint must be
placed on top of the one with the larger footprint at all times.
• The bottom device must be correctly levelled.
• Be sure to use the appropriate stacking adapter.
• The levelling feet on the top device must be screwed in all the way.
• The levelling feet of the top device must be aligned with and placed exactly on the stacking
pads of the stacking adapter.
• The anti-tilt anchor must be installed on the top device.
Thermo Scientific
1
1
1
Installing the Stacking Adapter
Anti-tilt anchor
Installation
Heratherm
3-11