Figure 10-1 Door Seal Replacement - Thermo Scientific IMH 60 Operating Instructions Manual

Heratherm advanced protocol and advanced protocol security microbiological incubators
Hide thumbs Also See for IMH 60:
Table of Contents

Advertisement

Maintenance
Temperature Calibration Procedure
Temperature Calibration Procedure
For detailed instructions on how to perform a manual temperature calibration, please refer to
the instructions in the section
Replacing the Door Seal
The door seal of the outer door is located in the retaining slot.
The door seal should be inspected for any signs of embrittlement at half-yearly intervals.
No tools are required to replace the seal.

Figure 10-1 Door Seal Replacement

1. Pull the seal out of the guide slot.
2. Starting on the hinge side of the door, position the end of the new seal at the location
indicated by the arrow in
3. Gently press the seal into the slot, working around the circumference of the door. In
corner areas in particular ensure that the seal lip is installed without any wrinkles and that
the seal is not stretched or compressed.
4. Make sure that the retaining rail taper is positioned correctly in the slot and that the seal is
flush with the door frame.
10-4
Heratherm
"Calibration"
Excessive work space temperatures subsequent to a calibration run
can be reduced by leaving the door open for approx. 30 seconds.
figure 10-1
on
page
7-23.
Excessive Work Space
Temperatures
above.
Thermo Scientific