Connecting The Rs-232 Interface - Thermo Scientific IMH 60 Operating Instructions Manual

Heratherm advanced protocol and advanced protocol security microbiological incubators
Hide thumbs Also See for IMH 60:
Table of Contents

Advertisement

Figure 5-10

Connecting the RS-232 Interface

The RS-232 data communication interface supports the querying of status information and
temperature data from the incubator by entering basic commands in a standard terminal
window provided by your computer's operating system. The interconnection requires a
standard RS-232 cable with 9-pin connectors and a straight "1:1" pinout without any crossed
wires, which is not supplied with the incubator.
Users may employ the RS-232 command inventory listed in
process data logging - for example, by embedding these commands in scripts that run on a
remote computer.
Thermo Scientific
AC Power Supply Socket
When taking the incubator into operation for the first time allow some
time before switching on for stabilization to avoid condensation forming
on live parts.
If a high-voltage test is to be performed on the unit, it must first be
heated for around 30 minutes at 75°C.
Replacement should only be carried out by skilled and authorized
qualified personnel of electrotechnology/signal engineering!
RS-232 interface check the interfacing parameters against the pin-out
description given above and make sure that computer's interface port
works with a signal level of +/- 5V DC.
Connecting the RS-232 Interface
Condensation
table 5-1
below for automating
RS-232 interface compatibility
To avoid overloading and damaging the
Installation procedures
Heratherm
5-9

Advertisement

Table of Contents
loading

Table of Contents