Calibrating The Shaker Speed - Thermo Scientific MaxQ SHKE416HP Operating And Maintenance Manual

High performance orbital shakers
Table of Contents

Advertisement

Section 2
Operation
Turning Audible Alarm
Off and On (cont.)
Calibrating the
Shaker Speed
2-10
Orbital Shakers
When the audible alarm is disabled, a
warning message is placed in the Setpoint
portion of the Operating Screen display as
illustrated at the right.
From the Operating screen, press the down arrow, up
arrow and Silence button in that sequence (shown at
the right) to open the configuration menu. From the
screen below, press the down arrow twice to bring up
the following screen.
Press the Speed button beneath RPM.
The value shown on the screen following is the present speed setpoint.
Using the up and down arrows, increase or decrease the platform speed
until the reading on an independent, accurate speed measuring device
matches the shaker speed setpoint.
When finished, press the Speed button to save the setting. The display will
return to the Calibrate - RPM Temp screen. Or, if nothing is pressed for
about fifteen seconds, the display will revert to the Operating Screen and
the setting will be automatically saved to memory.