Section 2
Operation
Setting the Remote
Alarms (continued)

Cycle Complete

Power Failure

2-12
Orbital Shakers
These alarms can alert a remote alarm monitoring system through an
internal SPDT relay connected to an RJ-11 jack on the rear of the shaker
cabinet. Refer also to "Connecting the Remote Alarm" in this manual.
Each of these alarms (except Platform Stalled and Check Fuse) may be
independently turned on and off to suit operator or laboratory needs.
To set the remote alarms to ON
or OFF, open the Remote Alarm
Configuration menu by pressing
the down arrow, up arrow, and
Silence buttons in that sequence.
Then, press the down arrow three times until the screen below is showing.
Press Rmte (remote). The alarms will be shown in the following sequence:
Actual
for more
Setpoints
RunHrs
Rmte
Note If no buttons are pressed, the display will automatically return to the
Operating Screen after about fifteen seconds, saving the selection to
memory.
s
Toggle the Cycle Complete alarm with either the up (ON) arrow or the
down (OFF) arrow. Pressing the Next (temperature) button advances the
display to the next alarm, saving the shown Cycle Complete setting to
memory.
Toggle the Power Failure alarm with either the up (ON) arrow or the down
(OFF) arrow. Pressing the Next (temperature) button advances the display
to the next alarm, saving the Power Failure setting to memory.