Calibrating Optional Temperature Probe; Setting The Remote Alarms - Thermo Scientific MaxQ SHKE416HP Operating And Maintenance Manual

High performance orbital shakers
Table of Contents

Advertisement

Calibrating Optional
Temperature Probe
Setting the Remote
Thermo Scientific
When using the optional temperature
monitoring probe, the system may be
calibrated +4°C from the factory setting.
Begin by pressing the down arrow, the
up arrow, the Silence button, and then
the down arrow twice.
The screen below will appear on the display.
Press the Temperature button beneath Temp.
Using the up and down arrows, increase or decrease the temperature value
to match an independent, accurate temperature measuring device. When
selected, press the Time, Speed, or Temp button to save the setting.
The display will return to the Calibrate - RPM Temp screen. If nothing is
pressed for about fifteen seconds, the display will revert to the Operating
Screen and the setting will be automatically saved to memory.
The SHKE416HP and SHKE430HP Orbital Shaker control system
monitors and provides alarms for six operating parameters:
Alarms
Parameter . . . . . . . . . . . . . . . . . . .Alarm Message
Cycle Status . . . . . . . . . . . . . . . . .Cycle Complete
Loss of Input Power . . . . . . . . . . . . .Power Failure
RPM versus Setpoint . . . . . . . . . . .RPM Tracking
Drive Belt Integrity . . . . . . . . . . . . . . .Check Belt
Platform Movement Status . . . . . .Platform Stalled
Motor Fuse Integrity . . . . . . . . . . . . . .Check Fuse
Section 2
Operation
Orbital Shakers
2-11