command:
:SENSE[n]:[CHANnel[n]]:POWer:WAVelength
syntax:
:SENSE[n]:[CHANnel[n]]:POWer:WAVelength<wsp>
<value>|MIN|MAX|DEF [PM|NM|UM|MM|M]
description:
Sets the sensor wavelength.
parameters:
The wavelength as a float value in meters.
Also allowed are:
response:
none
example:
sens1:pow:wav 1550nm
affects:
All power meter modules
command:
:SENSE[n]:[CHANnel[n]]:POWer:WAVelength?
syntax:
:SENSE[n]:[CHANnel[n]]:POWer:WAVelength?
[<wsp>MIN|MAX|DEF]
description:
Inquires the current sensor wavelength.
parameters:
none
Also allowed are:
response:
The wavelength as a float value in meters.
sens1:pow:wav? → +1.55000000E-006<END>
example
affects:
All power meter modules