Introduction to Programming
Measurement Functions – The SENSe Subsystem
Setting parameters for the stability function sets some parameters for
the logging and MinMax functions and vice versa.
:SENSe[n][:CHANnel[n]]:FUNCtion:PARameter:STABility
:SENSe[n][:CHANnel[n]]:FUNCtion:PARameter:STABility<wsp>
<total time>[NS|US|MS|S],<period time>[NS|US|MS|S],
<averaging time>[NS|US|MS|S]
Sets the total time, period time, and averaging time for the stability function.
Total time:
The total time from the start of stability mode until it is
completed.
Period time:
A new measurement is started after the completion of
every period time. Period time
Averaging time:
A measurement is averaged over the averaging time.
1
2
The total time should be longer than the period time.
The period time should be longer than the averaging time.
The number of data points is equal to the total time divided by the period
time.
Total time, period time, and averaging time are time values in seconds.
If you specify no units in your command, seconds are used as the default.
none
sens1:func:par:stab 1s,0.1s,0.1s
All power meter modules
82
Averaging Time
Period Time
3
4
5
t