command:
syntax:
description:
parameters:
response:
example:
affects:
NOT E
command:
syntax:
description:
parameters:
response:
example:
affects:
Introduction to Programming
Measurement Functions – The SENSe Subsystem
:SENSe[n][:CHANnel[n]]:FUNCtion:PARameter:LOGGing?
:SENSe[n][:CHANnel[n]]:FUNCtion:PARameter:LOGGing?
Returns the number of data points and the averaging time for the logging
function.
none
Returns the number of data points as an integer value and the averaging
time, t
, as a float value in seconds.
avg
sens1:func:par:logg?
All power meter modules
Setting parameters for the MinMax function sets some parameters for
the stability and logging functions and vice versa.
:SENSe[n][:CHANnel[n]]:FUNCtion:PARameter:MINMax
:SENSe[n][:CHANnel[n]]:FUNCtion:PARameter:MINMax<wsp>
CONTinous|WINDow|REFResh,<data points>
Sets the MinMax mode and the number of data points for the
MinMax function.
CONTinous:
WINDow:
REFResh:
Data Points is the number of samples that are recorded in the memory
buffer used by the WINDow and REFResh modes.
Data Points is an integer value.
See Chapter 3 of the HP 8163A Lightwave Multimeter & HP 8164A
Lightwave Measurement System User's Guide, for more information on
MinMax mode.
none
sens1:func:par:minm WIND,10
All power meter modules
80
+64,+1.00000000E-001<END>
continuous minmax mode
window minmax mode
refresh minmax mode