command: :STATus:QUEStionable:CONDition?
syntax:
:STATus:QUEStionable:CONDition?
description:
Returns the Questionable Status Condition Summary Register.
parameters:
none
response:
The sum of the results for the Questionable Status Condition Summary
Register as a 16-bit signed integer value (0 .. +32767)
Bit
5-15
4
3
2
1
0
stat:ques:cond? → +0<END>
example:
command:
:STATusn:QUEStionable:CONDition?
syntax:
:STATusn:QUEStionable:CONDition?
description:
Returns the Questionable Slot Status Condition Register for slot n.
parameters:
none
response:
The results for the individual slot events (a 16-bit signed integer value, where
0 ≤ value ≤ 32767):
Bit
8-15
7
6
5
4
3
2
1
0
Every nth bit is the summary of slot n.
stat1:ques:cond? → +0<END>
example:
Instrument Setup and Status
Status Reporting – The STATus Subsystem
Mnemonic
Not used
Summary of slot 4
Summary of slot 3
Summary of slot 2
Summary of slot 1
Summary of slot 0
Mnemonic
Not used
Slot n: Duty cycle is out of range
Slot n: ARA recommended
Slot n: Module is out of specification
Slot n: Module has not settled
Slot n: Laser protection on
Slot n: Temperature out of range
Slot n: Zeroing failed
Slot n: Excessive Value
62
Decimal Value
0
16
8
4
2
1
Decimal Value
0
128
64
32
16
8
4
2
1