HP 8163A Programming Manual page 128

Lightwave measurement system
Hide thumbs Also See for HP 8163A:
Table of Contents

Advertisement

command:
:TRIGger:CONFiguration?
syntax:
:TRIGger:CONFiguration?
description:
Returns the hardware trigger configuration.
parameters:
none
response:
DIS:
DEF:
PASS:
LOOP:
CUSTOM:
trig:conf? → DEF<END>
example:
Introduction to Programming
Triggering - The TRIGger Subsystem
Trigger connectors are disabled.
The Input Trigger Connector is activated, the incoming
trigger response for each slot
":TRIGger[n][:CHANnel[n]]:INPut" on page 123
determines how each slot responds to an incoming trigger, all
slot events (see ":TRIGger[n][:CHANnel[n]]:OUTPut" on
page 125) can trigger the Output Trigger Connector.
The same as DEFault but a trigger at the Input Trigger
Connector generates a trigger at the Output Trigger
Connector automatically.
The same as DEFault but a trigger at the Output Trigger
Connector generates a trigger at the Input Trigger Connector
automatically.
A custom configuration is active using either the command
":TRIGger:CONFiguration:EXTended" on page 129 or the
HP 816x VXIplug&play Instrument Driver. See Appendix A
"The HP 816x VXIplug&play Instrument Driver".
128