command:
[:SOURce[n]][:CHANnel[n]]:MODout
syntax:
[:SOURce[n]][:CHANnel[n]]:MODout<wsp>FRQ|FRQRDY|0|1
description:
Sets the modulation output
parameters:
FRQ or 0:
FRQRDY or 1:
response:
none
example:
sour2:mod 0
affects:
All Tunable Laser modules
command:
[:SOURce[n]][:CHANnel[n]]:MODout?
syntax:
[:SOURce[n]][:CHANnel[n]]:MODout?
description:
Returns the mode of the modulation output.
parameters:
none
response:
modulation signal is output all the time
0:
modulation is combined with the laser-ready signal.
1:
In this case, the output is kept low when no optical signal is output (for
example, while the laser is settling after a change of wavelength).
example:
sour2:mod?
affects:
All Tunable Laser modules
Introduction to Programming
Signal Generation – The SOURce Subsystem
modulation signal is output all the time
modulation is combined with the laser-ready signal.
In this case, the output is kept low when no optical signal
is output (for example, while the laser is settling after a
change of wavelength).