HP 8163A Programming Manual page 86

Lightwave measurement system
Hide thumbs Also See for HP 8163A:
Table of Contents

Advertisement

command:
:SENSe[n]:[CHANnel[n]]:POWer:RANGe[:UPPer]
syntax:
:SENSe[n]:[CHANnel[n]]:POWer:RANGe[:UPPer]<wsp> <value>[DBM]
description:
Sets the power range for the module.
The range changes at 10 dBm intervals. The corresponding ranges for linear
measurements (measurements in Watts) is given below:
+30 dBm
+20 dBm
+10 dBm
-10 dBm
-20 dBm
-30 dBm
-40 dBm
parameters:
The range as a float number in dBm. The number is rounded to the closest
multiple of 10, because the range changes at 10 dBm intervals. Units are in
dBm.
response:
none
example:
sens1:pow:rang -20DBM
affects:
All power meter modules
command:
:SENSe[n]:[CHANnel[n]]:POWer:RANGe[:UPPer]?
syntax:
:SENSe[n]:[CHANnel[n]]:POWer:RANGe[:UPPer]?
description:
Returns the range setting for the module
parameters:
none
response:
The range setting as a float value in dBm
(−110 ≤ value ≤ +30).
example:
sens1:pow:rang?
affects:
All power meter modules
Introduction to Programming
Measurement Functions – The SENSe Subsystem
Range
Upper Linear
Power Limit
1999.9 mW
199.99 mW
19.999 mW
1999.9 µW
0 dBm
199.99 µW
19.999 µW
1999.9 nW
199.99 nW
86
Range
-50 dBm
-60 dBm
-70 dBm
-80 dBm
-90 dBm
-100 dBm
-110 dBm
-2.00000000E+001<END>
Upper Linear
Power Limit
19.999 nW
1999.9 pW
199.99 pW
19.999 pW
1.999 pW
0.199 pW
0.019 pW