Detection Function
1
sensor, or the READY lamp on the docking stand is lit green. Even if the indicator is not lit green,
radiation can be delivered but an image will not be output. Make sure that the indicator is lit green
before making an exposure.
2 Check the tube current of the X-ray equipment in advance, and set exposure conditions based
on the tube current by referring to the table below. If the conditions are not met, X-rays cannot be
detected automatically and an image may not be acquired.
Tube current
Tube voltage
More than
Set the tube voltage
40 mA
according to the
anatomical region
and body thickness.
More than
Set the tube voltage
20 mA and less
to more than 50 kV
than 40 mA
according to the
anatomical region
and body thickness.
More than
10 mA and less
than 20 mA
Less than
The automatic X-ray detection function cannot be used.
10 mA
not limited to the bone region only.
subject, must have a width of more than 3 cm (1.2 in.) from the subject.
*3 Depending on the X-ray equipment, the actual exposure time may differ from the set time. Before use,
3 As illustrated below, if the subject whose thinnest part is at least 40 cm (15.7 in.) in thickness
image may not be acquired. In this case, make an exposure for the region including the directly
exposed area or switch to the High Sensitivity Mode. While the high sensitivity mode is being set,
a shock might cause a blank image.
Figure of an exposure for the subject covering the entire surface of the flat panel sensor
Exposure
SID
time
More than
Set the SID according to
1 ms
the anatomical region.
(*3)
Set the SID to 100 cm (39.4 in.) or less and do not
Alternatively, set the SID according to the anatomical
region and include the directly exposed area (*2).
100 cm (39.4 in.) or less
Do not limit the radiation
(*1) only.
Include the directly
exposed area (*2).
II
FDR D-EVO
Operation Manual
897N120339A
Z-7