System Parameters - GE Optima XR642 Pre-Installation

Table of Contents

Advertisement

GE Healthcare
Direction 5743854-1EN, Revision 1
most popular networking connection to ease integration into your facility's network.
In the system room, this facility;
Has 100baseT installed
Will have 100baseT installed
Do you segment your network using subnets?
Yes
Our equipment's IP addresses are:
Static
2.5

System Parameters

The system default uses the following IP Addresses internally:
192.168.1.1 eth0
192.168.2.1 eth1
192.168.3.1 br0 (eth2/3)
192.168.1.30 detector IP before detector boot
192.168.1.50 detector IP after boot in table
192.168.2.45 detector IP after boot in WS
192.168.3.50 detector IP after boot in Tether/wireless
If the hospital network uses 192.168.x.x, there will be a conflict. If this conflict occurs, you must
contact your GE Service Representative to change the internal IP Addresses used by the system.
Has 10baseT installed
Will have 10baseT installed
Acquired via DHCP
System
Host Name: _________________________
Network (IP) Address: _ _ _ . _ _ _ . _ _ _ . _ _ _
Subnet Mask: _ _ _ . _ _ _ . _ _ _ . _ _ _
Router IP: _ _ _ . _ _ _ . _ _ _ . _ _ _
Scheduled Station AE Title: _________________________
Chapter 6 - Communications/Networking
Optima XR642/XR648 Pre-Installation
No
Has a different network installed
We don't have a network installed
A combination of both methods
Page 113