GE Optima XR642 Pre-Installation page 83

Table of Contents

Advertisement

GE Healthcare
Direction 5743854-1EN, Revision 1
Illustration 2-41 : Standard Table & Manual WS at Head with 3M bridge
Item
Limit
Dimension
mm
inch
(1)
Min
630
24.8
(2)
Min
203
(3)
Min
2060
81.1
Max
2735
107.7
Recommend
Dimension
mm
inch
700
27.6 Head end of rail to center line of WS detector.
With manual WS, the Min value should be larger than 630 mm
(530+100 Cal space)
8.0
400
15.7 Service clearance for WS installation.
The Min value should be larger than 203 mm, the WS back cover
could be installed or can be removed.
2300
90.6 Back of bridge to center line of WS.
1.With manual WS, the Min value should be larger than 2060
(860 +100 Cal space + 700 distance between focal spot to back of
OTS column +400 half table Width)
2.with manual WS, the Max value should be less than 2735 ( OTS
biggest travel range2835 - 100 Cal space)
Chapter 2 - Equipment
Optima XR642/XR648 Pre-Installation
Description
Page 83