GE Optima XR642 Pre-Installation page 89

Table of Contents

Advertisement

GE Healthcare
Direction 5743854-1EN, Revision 1
Room size summary for full service function summary
Configuration
Standard Table + Manual WS or
G2 Standard  WS(L*W*H)
Room type
Min (mm)
w/o control room
4750*3341*2686
w/o stretcher
table
w/o control room
4750*4000*2686
w stretcher table
w control room
4750*4800*2686
Note: G2 Extended WS: Min_Ceiling Height=2750mm
Min Room size for limited service function (special case)
Need Service FE and customer accept below 2 limitations:
1.) Tabletop can't be pulled out in exam room, so table actuator and tabletop can't be replaced in exam room.
(Need take table out of the room if have to pull overall tabletop out for troubleshooting)
2.) FE need install tabletop out of exam room, then push table into exam room.
Then system could be installed into the site with below 2 merits and above 2 limitations:
1.) Min room size: 4217*3677mm.
2.) Tabletop could be pushed to head/foot side until 300mm table frame is exposed, then the bolt of top rear cover
could be exposed. Remove all table cover. most components could be repaired except replacing table actuator
and tabletop.
Standard Table + Extended
WS(L*W*H)
Recommend
Min (mm)
(mm)
5250*4000*2986
4750*3341*2750
5250*5000*2986
4750*4000*2750
5250*5000*2986
4750*4800*2750
Configuration
Room type
w/o control room 
w/o stretcher table
w/o control room 
w stretcher table
w control room
Chapter 2 - Equipment
Optima XR642/XR648 Pre-Installation
Manual WS only(L*W*H)
Recommend
Min (mm)
(mm)
5250*4000*2986
3800*3000*2686
5250*5000*2986
3800*3000*2686
5250*5000*2986
3800*3500*2686
Standard Table + Manual WS
or G2 Standard  WS(L*W*H)
Min (mm)
4217*3341*2686
4217*4000*2686
4217*4800*2686
Recommend
(mm)
4000*3500*2986
4000*3500*2986
4000*4000*2986