GE Optima XR642 Pre-Installation page 84

Table of Contents

Advertisement

GE Healthcare
Direction 5743854-1EN, Revision 1
(4)
Min
960
(5)
Min
1100
(6)
Min
102
(7)
Min
62
(8)
Fix
3077
121.1
(9)
Min
100
(10)
Min
50
(11)
Fix
1660
(12)
Fix
660
(13)
Fix
2330
(14)
Min
50
(15)
Min
1353
(16)
Min
1307
(17)
Min
200
Page 84
37.8
1000
39.4 Back of bridge to center line of table.
With manual WS, the Min value should be larger than 1960 (860
+100 Cal space + 600 distance between focal spot to back of OTS
column +400 half table Width)
43.3
1300
51.2 Table center line to WS center line.
The Min value should be larger than 1000 mm (700 distance
between focal spot to back of OTS column + 400 half table width)
4.01
200
7.9 Room wall to drape rail
Make sure cable not touch the wall
2.44
200
7.9 Drape rail to back of bridge.
Make sure cable could bend freely
3077
121.1 3M Bridge length
3.94
500
19.7 Service clearance for OTS installation.
2.0
100
3.9 Service clearance for Tabletop at head limit postion.
The min value for tabletop not touch the wall.
65.4
1660
65.4 Half length of tabletop longitudinal travel coverage
26.0
660
26.0 Half length of table base (no tabletop)
91.7
2330
91.7 Length of tabletop
2.0
500
19.7 Service clearance for Tabletop pulled out.
The min value for tabletop not touch the wall when it is pulled
out. FE could stand by side of the tabletop if room length is short.
53.3
1400
55.1 Service clearance for cabinet.
51.5
1400
55.1 Service clearance for cabinet.
7.9
200
7.9 Service clearance for cabinet.
Section 3.0 - Room Layout
Optima XR642/XR648 Pre-Installation

Advertisement

Table of Contents
loading

This manual is also suitable for:

Optima xr648

Table of Contents