Iso-Channel Eliminates Ground Loops And Increases Common-Mode Rejection - Data Translation TEMPpoint DT9871U User Manual

Usb measurement instrument
Hide thumbs Also See for TEMPpoint DT9871U:
Table of Contents

Advertisement

Appendix C
ISO-Channel Eliminates Ground Loops and Increases
Common-Mode Rejection
ISO-Channel technology eliminates ground loop problems by using a differential, isolated,
floating front-end. As you can see in
system ground.
To measure floating signal sources, ISO-Channel technology uses differential analog input
signals, a 24-bit Delta-Sigma A/D converter for each channel, and channel-to-channel
isolation, as shown in
126
Figure
37, a floating voltage signal is not referenced to
Figure 37: Floating Signals
Figure
38.