Memory Specifications
Table 14
lists the memory specifications for the analog input subsystem on the TEMPpoint,
VOLTpoint, and MEASURpoint instruments.
Data memory onboard
For Data logger built in, maximum time
before old data is overwritten
48 channels @ 10 Hz:
48 channels @ 1 Hz:
48 channels @ 0.1 Hz:
a. Assumes limit detection is off for all channels, and for thermocouple channels,
assumes CJC data is not collected. If power fails, all temperature data in the
system is lost. The channel input type and filter settings are still available after
power on, but the channel and digital I/O labels and channel limits are lost.
Table 14: Memory Specifications
Feature
a
Specifications
4 MByte
30 minutes
5 hours
50 hours
Specifications