TROUBLESHOOTING

In the event of faults that are not listed in this chapter, or if the
suggested procedure does not correct the fault, contact the
PARI GmbH Service Center.
Fault
The
compressor
does not start
up.
No medication
is coming out
of the
nebuliser.
– 77 –
Possible cause
The power cord is not
plugged into the device
properly.
The power plug is not
plugged into the mains
socket properly.
The nebuliser nozzle is
blocked.
The compressed air
supply tube is not
connected to the air
connection on the
compressor correctly.
The tube attached to the
air connection on the
compressor is leaking.
Procedure
Check that the power
cord is seated correctly
in the compressor
power cord socket.
Check that the power
plug is seated correctly
in the mains socket.
Clean the nebuliser.
Check that both ends of
the tube are seated
properly.
Replace the defective
tube or the entire
PARI SINUS tubing
system.