2 Installing the Bravo Platform
Starting up and shutting down
Shutting down the Bravo Platform
C AU T I O N
Related information
52
Shut down the Bravo Platform before you:
Clean the Bravo Platform
Change the liquid-handling head
Install accessories
Move the Bravo Platform
Always turn off the Bravo Platform before installing or uninstalling a liquid-handling
head. Failure to do so can damage the head electronics.
To shut down the Bravo Platform:
1
If removing the liquid-handling head, ensure that the disposable tips are removed.
2
Optionally, home the liquid-handling head.
3
Shut down the computer.
4
Turn off any accessories, for example, the Pump Module.
5
If using an Auto Filling Reservoir, disconnect the bottles to prevent siphoning.
6
On the side of the Bravo Platform, press the power switch to the off (o) position.
7
If applicable, uninstall the liquid-handling head.
For information about...
Setting up and using the Bravo
Platform
G5562A, G5563A Bravo Platform Safety and Installation Guide