2 Installing the Bravo Platform

Starting up and shutting down

Removing the dust cover

Starting up and shutting down

About this topic
Starting up the Bravo Platform
C AU T I O N
50
To remove the dust cover:
1
Ensure that the Bravo power switch is set to the off (o) position.
2
Lift the dust cover up and off of the Bravo Platform.
This topic describes how to start up and shut down the Bravo Platform when you are
operating it as a standalone device. For instructions on how to turn on and turn off the
Bravo Platform when it is integrated into a workstation or system, see the workstation
or system user documentation.
Always turn off the Bravo Platform before installing or uninstalling a liquid-handling
head. Failure to do so can damage the head electronics.
To start up the Bravo Platform:
1
Ensure that the main power cable and Ethernet are plugged into the connection
panel.
2
Turn on any accessories, for example, Pump Module.
3
Turn on the computer and the monitor, and start the Microsoft Windows operating
system.
4
If you have not already done so, install the liquid-handling head.