1 Safety guidelines

Emergency stops

Emergency stops
WAR NING
C AU T I O N
Stopping in an emergency
Recovering from an emergency stop
6
If the Bravo Platform is integrated with other devices in a system, Agilent
Technologies recommends that you install a main emergency-stop unit, such as the
Robot Disable Hub, that will stop all devices simultaneously. In addition, all operators
must be instructed in the emergency-stop procedure.
You might not be able to resume a protocol after an emergency stop. Do not use an
emergency stop to pause a run. To pause and continue a run, use the appropriate
commands in the VWorks software.
To stop in an emergency:
Press the red button on the emergency-stop pendant. The Bravo head stops
immediately.
Figure
Emergency-stop pendant
To recover from an emergency stop
1
Remove any labware that was dropped and clean up any spills.
2
Release the red button on the pendant by turning it clockwise.
3
In the VWorks software message box, select one of the options to abort, retry, or
continue. This will re-enable the axes motors.
G5562A, G5563A Bravo Platform Safety and Installation Guide