G5562A, G5563A Bravo Platform Safety and Installation Guide
Note: If you do not hear the pins snap into place, check that the straight edges
of the retainer pins are in the vertical position, as the following figure shows.
Attempt to rotate the pins to ensure that they are in the locked position. The
pins should not rotate freely.
Figure
Installed Bravo 96AM Head: (
2
7
To lock the head, rotate the head lock clockwise until it reaches its hard stop. This
ensures that the head is fully seated and does not shift position during operation.
Dropping the head or bumping the barrels or probes will damage the head. If the
Bravo head is not properly secured in place, it could drop unexpectedly. Ensure that
the head is securely locked into the head mount.
2 Installing the Bravo Platform
Installing the liquid-handling head
1
2
) retainer pin and (
) head lock