2 Installing the Bravo Platform
Installing the Standard Light Curtain and shields
2
Aligning the light beams
40
To connect the Standard Light Curtain using the Robot Disable Hub:
Refer to the following figure and table for the connections.
Figure
Connections for Standard Light Curtain and pendant
1
Pendants
Power
Devices
Item
Description
1
Pendant cable
2
Jumper
3
Extension cable
4
Lightpost cable,
black
5
Lightpost cable, gray
To align the light beams:
1
Ensure that the Light Curtain switch on the Robot Disable Hub is set to Enable.
2
On the side of the Bravo Platform, press the power switch to the on (|) position.
The two indicator lights on the front of the Bravo backplate illuminate, indicating
that the Bravo Platform is on, and the Standard Light Curtain activates.
G5562A, G5563A Bravo Platform Safety and Installation Guide
4
Rx
Light Curtain
5
Tx
3
From...
Pendant
Jumper
Bravo Pendant port
Lightpost receiver
Lightpost transceiver
00220
Bravo
Back
To Robot Disable Hub
Pendants connector
Pendants connectors
Devices connector
RX port
TX port