Agilent Technologies G5562A Safety And Installation Manual page 30

Bravo platform
Hide thumbs Also See for G5562A:
Table of Contents

Advertisement

2 Installing the Bravo Platform
Connecting the Bravo Platform
Pump I/O port
22
Item
Feature
2
Pendant port
3
Fuse holder
4
AC power entry
The following figure shows the Pump I/O port on the back of the Bravo Platform. The
table below describes the Pump I/O port in detail.
Figure
Pump I/O port (Bravo Platform back view)
00212c
Bravo
Acc_IO
Item
Feature
1
Pump I/O port
G5562A, G5563A Bravo Platform Safety and Installation Guide
Description
Connects the emergency-stop pendant and a Bravo
light curtain to the Bravo safety interlock circuit.
The connection is made through an emergency-stop
control, such as the Robot Disable Hub.
IMPORTANT
The safety interlock circuit must be
closed for the Bravo Platform to operate.
Houses the AC inlet fuse and a spare fuse for the Bravo
Platform.
Connects the Bravo power cord to an AC outlet with a
grounded circuit.
1
Description
An RJ-45 I/O port that connects the Pump Module to
the Bravo Platform. The connection is made with a
straight-through shielded Cat-5 or Cat-6 (Ethernet)
cable.
IMPORTANT
This is not an Ethernet port.