2 Installing the Bravo Platform
Installing the Standard Light Curtain and shields
Installing the side shields
44
Figure
Bravo Platform front shield attachment screws
Bravo
The Bravo side shield configurations can vary depending on whether the Bravo Platform
is integrated with another device at a given side. The options include:
Full-size side shield. Prevents access to the Bravo moving parts from the side. The
full-size shield is used on a side where no device is integrated.
Upper side shield. Prevents access to the Bravo moving parts at the upper side only.
For example, the upper shield could prevent an operator from reaching over a
device with a low profile that is integrated at the side of the Bravo Platform.
Partial-width shield. Prevents access from the side to certain deck locations only.
For example, if a device is integrated at the side aligned with the back row of
platepads, a partial side shield can prevent access to the other rows of platepads
that are not blocked by an integrated device.
The procedure in this section describes how to install the full-size side shield. Refer to
the following figure and table for this procedure.