The Yokogawa ZA8C and AV8C Convertor Electronics setup

The DR probe can be wired to work with the Yokogawa® ZA8C and AV8C Converters. Con-
nect the cabling from the ZA8C or AV8C terminal to the probe terminal in the junction box
as shown in Figure 2-16.
Figure 2-16. DR Probe Wired to the ZA8C or A V8C Converter
THERMOCOUPLE
JUNCTION
GND
32
SHIELD
HEATER
ZA8C OR AV8C
ZA8C
OR
AV8C
CONVERTER
Notes:
1. Heater temperature set to 1380°F (750°C)
2. The greater mass of the oxymitter direct replacement probe requires lohnger time to heat up. Upon
startup, the Yokogawa electronics may indicate an error because the probe has not reached temperature
setpoint in the normal time. Remove power from the Yokogawa electronics or probe module to clear
the error, and restore power. This procedure may have to be repeated a couple of times before operating
temperature is reached.
52
ZA8C
OR
AV8C
TERMINAL
}
3+
CELL
4–
}
5+
6–
}
7+
COLD
8–
16
PROBE
17
HEATER AND SIGNAL CABLE
OXYMITTER DR PROBE
WITH ADAPTER FLANGE
+
STACK WALL
EXISTING
MOUNTING
July 2018
COLD
JUNCTION
DEVICE
SHIELD
36430002
Installation