Sps 4001B And Imps 4000, 1-4 Probes; Additional Imps 4000 Alarm Contacts; Identifying And Correcting Alarm Indications - Emerson Rosemount Oxymitter 4000 Reference Manual

Hazardous area oxygen transmitter
Hide thumbs Also See for Rosemount Oxymitter 4000:
Table of Contents

Advertisement

Rosemount Oxymitter 4000

SPS 4001B and IMPS 4000, 1-4 probes

Additional IMPS 4000 Alarm Contacts

Identifying and Correcting Alarm Indications

98
One contact closure per probe from the control room to the SPS 4001B or IMPS
4000 for "calibration initiate".
One contact output per probe from the SPS 4001B or IMPS 4000 to the control
room for "in calibration" notification.
One contact output per probe from the SPS 4001B or IMPS 4000 to the control
room for "calibration failed" notification. (Includes output from pressure switch
indicating "cal gas bottles empty").
One contact per IMPS 4000 for "low calibration gas flowing".
One contact per IMPS 4000 for "high calibration gas flowing".
NOTE
The 4-20 mA signal can be configured to respond normally during any calibration, or can
be configured to hold the last O
is for the 4-20 mA signal to operate normally throughout calibration.
NOTE
Holding the last O
value may be useful if several probes are being averaged for the
2
purpose of automatic control. Unless several probes are being averaged, always place
any control loops using the O
For a Hazardous Area Oxymitter 4000 with a membrane keypad, faults are indicated by
four diagnostic, or unit, alarm LEDs. A pattern of repeating blinks define the problem. A
condensed table of the errors and the corresponding blink codes can be found on the
inside right cover of the electronics housing. Table 8-1 also identifies the blink code and
fault status of each LED as well as the output of the 4-20 mA signal line and a fault
number that corresponds to the troubleshooting instructions provided in this section.
For a Hazardous Area Oxymitter 4000 with the optional LOI, alarm messages are
displayed on the LOI display window when the alarm status display is accessed via the
LOI menu. A listing of the alarm/fault messages and the related fault status descriptions
and fault numbers are shown in Table 8-2.
value upon the initiation of calibration. Factory default
2
signal into manual prior to calibrating.
2
July 2018
Troubleshooting

Advertisement

Table of Contents
loading

Table of Contents